SET-1215

RTD-SIM (PT100/PT1000) Card

High-precision resistance thermometer simulator

Übersicht SLSC Cards
SET-1215 RTD simulation card

SET 1215 – RTD-SIM (PT100/PT1000)

The SET-1215 RTD simulation card is a high precision 16 channel resistance temperature detector simulator.
The simulation range of the SET-1215 RTD simulation card is from 90 Ohm to 410 Ohm and 900 Ohm to 4100 Ohm with an accuracy up to 1°C.

It contains 16 galvanic isolated channels which can be connected in 2-terminal sensing and 4-terminal sensing.
Each channel of the SET-1215 card is able to simulate the common short circuit and open circuit conditions, which could be experienced in a system due to faulty wiring or sensors.

To maximize customizability, the SET-1215 have two plug-in module slots that can provide features like line fault insertion and instrument connect. Additionally, the SET-1215 RTD Card comes with a high current fault injection bus.

Artikel anfragen

SET-1215

SKU:
104091

  • HiL Test Systems
  • Automotive
  • Aerospace

Absolute Maximum Ratings

ConditionValueComment
Max. Input VoltageAny Pin to Chassis GND 60 VDCTransient 65V, Limited by connector
Min. Input VoltageAny Pin to Chassis GND-60 VDC
Max. Input VoltageBetween CHx_F+ and ChxF-15 V
Min. Input VoltageBetween CHx_F+ and ChxF--0.3 V
Max. Short Circuit Current1.5 A
Max. Simulation Current10 mA
Min. Simulation Current100 µA
Max. Switching PowerDC, Resistive load60 W
Max. Simulation Power400 mW

Technical Data

ConditionValueComment
Max. Simulation Voltage Between CHx_F+ and ChxF-10 V
Min. Simulation VoltageBetween CHx_F+ and ChxF-0 V
Resistor Value Update Time 50 µsSLSC communication delay not included
Simulation ResolutionPT100 configuration PT1000 configuration
Initial Resistance< 10 M
Expected Electrical Relay Lifetime
PT100 Simulation Range min.90
PT100 Simulation Range max.400
PT1000 Simulation Range min.900
PT1000 Simulation Range max.4000

Physicals Characteristics

ConditionValueComment
Module DimensionsExcluding ejector144.32mm x 30.48mm x 302mm (H x W x D)Standard SLSC card size
Front Panel Connector2x female DB -44 high-density D-Sub with 4-40 UNC screw lockFor mating connectors and cables, contact us
RTI Connector 2mm hard metric per IEC 61076-101Any RTI marked

Environmental

ConditionValue
Operating Humidity Relative, non-condensing10% – 90%
Storage HumidityRelative, non-condensing5% – 95%
Operating TemperatureForced-air cooling from chassis0 °C – 40 °C
Storage Temperature-40 °C – 85 °C
Maximum Altitude2000 m

SET-1215 Technical Description Download Documenation